HypotULTRA III / S7600 RS-232 Trouble Shooting Guide
Error 1 occured at Error with command "ADD..." - Event Register = 0: Remote Scanner Set Up as Local Scanner Background: The HypotULTRA III is capable of interfacing with both an internal and external Scanner controlled locally by the Hipot tester, or with a remote Scanner which is controlled by the Autoware software through a direct connection to the PC. When incorporating a Scanner into your testing procedure, it is important to define whether Autoware should treat the Scanner as a local Scanner or a remote Scanner. If you have selected that your Scanner is to be controlled locally and the Scanner is actually a remote Scanner that is controlled through Autoware directly, Autoware will display an error message similar to the one above (depending on the scanner configuration).
Solution: In order to solve this problem you will need to reconfigure your Scanner settings.
- From the Main Menu select the Setup System Parameters Menu.
- The first thing to do is disable the internal Scanner configuration. The fourth field from the top edits the Scanner parameters.
- Using the mouse, click on the up and down arrows to scroll through the different Scanner configurations until you have set the parameter to "None."
- Next you must configure the remote Scanner parameter to match your current configuration.
- Click the hot button labeled "Remote Scanners" under the More Settings heading at the bottom of the screen.
- Using the mouse, scroll through the arrows next to the heading "Number of Remote Scanners." Depending on you configuration, you can set up a total of 12 remote Scanners. Once you have selected the correct number of remote Scanners you must then edit each Scanner's operating parameters.
- Make sure that the "Scanner Type" matches your Scanner (SC6540 or HS-16).
- Set you communication type to either Serial (RS-232) or GPIB.
- Next configure the proper port/address for the current Scanner (for more information on this read the Background section under the Error - 1073807343). Finally, set the correct number of HV (high voltage) channels you have opted for with your Scanner.
- Once all of the parameters for each Scanner have been set correctly click OK to return to the Setup System Parameters Menu.
- Click the OK button to return to the Main Menu.
|