Autoware S8456 Trouble Shooting Guide for RS-232 Interface
Error #1 - Error with command "FL1" - Event Register = 0: Incorrect DUT HV setup
Background: One of the features of OMNIA models 8105 and 8106 is the addition of a DUT output jack used to supply power to a DUT during Line Leakage and Run testing. This feature can be enabled from both the front panel of the OMNIA and the Autoware software. If there is a discrepancy between how the parameter is set in OMNIA's hardware and Autoware's software an error message will be displayed when attempting to save a file to the OMNIA.
Solution: In order to solve this problem you will need to make sure that the DUT-HV parameter is consistent in both OMNIA's hardware and Autoware's software.
- From the startup screen select Setup System Parameters located in the Window Menu.
- Under the Primary Unit tab, make sure that the box next to DUT-HV Setting is checked.
- Select Save and Close located in the File Menu to return to the startup screen.
- Exit the Autoware program by selecting Exit Autoware from the File Menu.
- From the Main Menu on the OMNIA's LCD screen, press the softkey corresponding to Setup System.
- From the Setup System Menu press the softkey corresponding to Hardware.
- Toggle the DUT-HV Setup parameter until it is set to ON.
- Press the Exit key twice to return to the Main Menu.
- Now that both the hardware and software DUT-HV Setting parameters are configured properly you may run the Autoware software normally.
|