Autoware S8456 Trouble Shooting Guide for RS-232 Interface
Error - 1073807360: Incorrect COM Port (Scanner/OMNIA)
*NOTE: This error can occur due to an incorrect Scanner configuration or an incorrect OMNIA configuration. If an OMNIA configuration error is suspected, follow the steps outlined in Error – 1073807343 to solve the problem. If a Scanner configuration error is suspected, follow the steps below.
Background: Automating your testing procedure requires that you interface a computer with your remote Scanner (a scanner controlled through a direct connection to the PC). In order to do this you can use the Autoware software to communicate with the remote Scanner in one of two ways: via an RS-232 interface or via a GPIB interface. If you have selected to use an RS-232 interface, your computer uses a serial port, which it labels "COM" followed by a number from 0-x (i.e. COM1). If Autoware tries to communicate with the remote Scanner using the wrong COM port you will receive an error message.
Solution: In order to fix this problem you need to make sure that you have identified the correct COM port so that Autoware can talk successfully to the Scanner.
- From the startup screen select Setup System Parameters located in the Window Menu.
- Click the Remote Scanners tab at the top of the screen.
- For each Remote Scanner being used, make sure that the Communication Type heading is set to "Serial."
- Using the mouse, scroll through the arrows beneath the Address heading until the correct port address is chosen for each Remote Scanner.
- Make sure that the number of HV (high voltage) and GND (ground bond) channels are correct for each Remote Scanner.
- Select Save and Close located in the File Menu to return to the startup screen.
|