Autoware S8456 Trouble Shooting Guide for RS-232 Interface
Error - 1073807195: Incorrect Communication Protocol Selected Background: Automating your testing procedure requires that you interface a computer with your OMNIA. In order to do this you can use the Autoware software to communicate with the OMNIA in one of two ways: via an RS-232 interface or via a GPIB interface. If you are using an RS-232 interface and you have configured Autoware to use a GPIB interface you will receive an error message.
Solution: In order to fix this problem you need to make sure that you have selected the Serial (RS-232) communication option. This selection allows Autoware to talk successfully to the OMNIA.
- From the startup screen select Setup System Parameters located in the Window Menu.
- Under the Primary Unit tab, the first field asks you to select the communications protocol to use. Make sure that this parameter is set to "Serial."
- The next field asks you to select which number port to use. Make sure that you select the number that corresponds to the port you have the RS-232 cable plugged into. *NOTE: You can click the View Available Communications Connections button to determine which COM port is being used.
- Select Save and Close located in the File Menu to return to the startup screen.
|